24
the dimming mode function.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, the soft-key zone
displays the function menu of this parameter. This parameter can be changed to OFF, Front, or BACK.
Timer
1) Description
This parameter is used to set the output timing function. When the user sets the value of this
parameter, the system will automatically shut down the output as the value countdown to 00:00:00.
If the user does not set this parameter value, the time of the instrument output is displayed under
the <MeasDisp> page.
2) Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it with
the front panel numeric keys, and finish with ENTER or OK keys to confirm the change. If you do not
want to change the parameter value or press the wrong numeric key when changing with the
numeric keys, you can cancel it by pressing the
key or ESC key.
VoltLiLmt
1)
Description
This parameter is used to set the output voltage limit protection. This parameter is the absolute value
of the difference between the output voltage and the set voltage. When the difference between real
output voltage and the set voltage is lower or higher than VoltLiLmt, the output initiates over-voltage
protection (OVP) or over-current protection (LVP) and the output is turned off.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it with
the front panel numeric keys, and finish with ENTER or OK keys to confirm the change. If you do not
want to change the parameter value or press the wrong numeric key when changing with the numeric
keys, you can cancel it by pressing the
key or ESC key.
3.3
<Programmable Mode>
3.3.1
Sequence Setting
3.3.1.1
Description and Operation
1) Description
In <ProgramSetup> MEMORY page, users can set the following parameters when the Surge/drop option
in <ProgramSetup> COMMON page is turned on (figure 3-10, 3-11, and 3-12 below).