167
Resolution
1°
Table 7-3 Instrument Setup
7.1.4
Measurement
7.1.5
Convention
Memory Sequence
50 groups
,
9 steps/group
Remote Signal Input
Output on, off, read manual mode memory sequence 1 to memory sequence
7
Remote Signal Output
Pass
、
Fail
、
Testing
Synchronous output signal
Output signal 10V
、
BNC type
Model
TH7105
TH7110
TH7120
Voltage
Range
0
~
300Vac
Resolution
0.1V
Accuracy
±(0.5% readings
+
2 digits)
Frequency
Range
45.0Hz
~
500Hz
Resolution
0.1Hz(45.0Hz
~
99.9Hz)
;
1Hz(100Hz
~
500Hz)
Accuracy
±0.1Hz
Current
Range
0
~
150Vac
0.000
~
4.200A
0.000
~
8.400A
0.000
~
16.800A
0
~
300Vac
0.000
~
2.100A
0.000
~
4.200A
0.000
~
8.400A
Resolution
0.001A
Accuracy
±(0.5% readings
+
3 digits)
AP
Range
0
~
150Vac
0.00
~
16.80A
0.00
~
33.60A
0.00
~
67.20A
0
~
300Vac
0.00
~
8.40A
0.00
~
16.80A
0.00
~
33.60A
Resolution
0.01A
Accuracy
±(5% readings
+
2 digits)
Power
Range
0~500W
0~1000W
0~2000W
Resolution
0.1W(0~1000W)
0.1W(0~1000W);
1W(1000~2000W)
Accuracy
±(0.6% readings
+
5digits) when PF>0.5
PF
Resolution
0.001
Accuracy
Measure and display to 3 digits significant figures