7
•
Maintenance and Service
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 181
7.11
Replacing the Drain Pump Head with Tubing
When
•
Drain pump tubing is worn or damaged
•
Drain pump head is broken
Parts and additional items required
Replacement drain pump head with tubing
Preparations
Make sure that the drain pump is turned off.
Follow these steps
Figure 81: Configuration of drain pump head with tubing (waste tubing)
No.
Description
No.
Description
1
Drain Pump Head
4
Straight Tubing Connector
2
Elbow tubing connector
5
Waste Tubing (pre-installed with
the pump head)
3
Condensation Tubing
1. Rotate the drain pump head counterclockwise slightly with and
remove it gently.
2. Disconnect the condensation tubing on the left from the tubing
connectors. Leave the tubing connectors connected to the drain
pump head.