Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 77
No.
Description
Shipped with
6
Depending on flow rate
Flush buffer loop for up to 5 mL/min, 50
µL, PEEK
Integral Fraction Collector FT
Flush buffer loop for 5 - 10 mL/min, 100 µL,
PEEK
Optional, needs be ordered separately
7
Waste fluidic
Detector
5.6.3.3
Connecting Viper Capillaries
This section describes how to connect
Viper™
capillaries. All Viper flow
connections in the Vanquish system are designed to be finger-tight.
To connect Viper capillaries with knurl, follow these steps:
NOTICE
•
Tighten or loosen Viper capillaries
only
with your fingers. Do not use
tools other than the knurl that comes with the capillary.
•
To avoid damage to the capillary or connection, tighten and loosen
the Viper capillaries only when the system pressure is down to zero.
Figure 26: Viper fitting with knurl
No.
Description
1
Knurl
2
Capillary
3
Slot
1. Insert the Viper capillary into the connection port.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...