7
•
Maintenance and Service
Page 170
Integral Fraction Collector FT (VF-F20-A) Operating Manual
a) Unpack the replacement capillary. Remove the protection cap
from the Viper fitting.
b) Connect the capillary to the flush pump by tightening the fitting.
c) Connect the Viper fitting to port 8 of fraction collection valve.
d) Remove the knurl from the Viper fitting due to the limited space
at the fraction collection valve.
NOTICE
A black Viper knurled head is needed to tighten or loosen Viper fitting.
Tighten the capillaries by hand. For details, see
(
page 77).
Keep the Viper knurls for reuse in the future.
11. Connect a new flush solvent tubing to port 7 of fraction collection
valve and route to the flush solvent reservoir in the solvent rack.
Figure 70:
Install the replacement flush solvent tubing
a) Unpack a replacement flush solvent tubing.
b) Connect the flush solvent tubing onto port 7 of fraction
collection valve by tightening the fitting.
c) Route the flush solvent tubing to flush solvent reservoir. For
Connecting Flush Solvent Reservoir (
page 86).
12. Power on the device. The device will run initialization automatically.
Make sure no errors or warnings are reported in the Audit trail.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...