7
•
Maintenance and Service
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 187
7.14
Replacing the Main Power Fuses
When
Blown fuses
Parts required
Replacement fuses (2 fuses, 5AT, 230 V AC, slow-blow, 5 x 20 mm) from
Fuses Kit
Tools required
Slotted screwdriver, any size between 3.3 mm and 5.5 mm is
appropriate
Preparations
WARNING
—
Electric Shock
High voltages are present inside the device that could cause an electric
shock or damage to the device.
Turn off the device with its main power switch. Disconnect the power
cord from both the power source and the device.
Use only the fuses of the type and current rating specified for the device
by Thermo Fisher Scientific. Do not use repaired fuses and do not short-
circuit the fuse holders.
Follow these steps
The fuse holder is located next to the main power switch.
Figure 83: Fuse holder