Integral Fraction Collector FT (VF-F20-A) Operating Manual
Parameters
Description
Collection path mode
Defines how the needle moves across the
racks.
Vertical
: The needle follows a vertical
meander-shaped path.
Saw Vertical
: The needle follows a vertical
saw-shaped path.
Horizontal
: The needle follows a horizontal
meander-shaped path.
Saw Horizontal
(default): The needle follows
a horizontal saw-shaped path.
Collection valve mode
Defines whether the fraction collection valve
stays at collection position or switches to
waste position when the needle moves from
one vial to next vial.
Interrupt
(default): In this mode, the fraction
collection valve switches to waste position
when the needle moves to the next vial,
during which the flow is directed to waste
container.
Continue
: In this mode, the fraction collection
valve remains at collection position when the
needle moves from vial to vial, during which
the flow keeps spraying out from the needle.
Needle positioning mode
This property refers to the pusher position
and needle height during fraction collection.
Different parameters as explained in the next
rows can be set to change the needle height
in different needle positioning mode (In Vial
or Above Vial) in Instrument Method Wizard
or Editor.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...