Maintenance and Service
Page 198
Integral Fraction Collector FT (VF-F20-A) Operating Manual
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to remove the slide-in module properly from the
enclosure in the next step.
3. Grasp the slide-in module by the middle frame where the door lock
is placed and pull the module out of the enclosure by approximately
10 cm.
Figure 86: Pulling out the slide-in module
NOTICE
The slide-in module can fall down when pulling it out of the
enclosure too far. Pull out the slide-in module just far enough so
that you can grasp it on both sides from below.
4. Remove the slide-in module from the enclosure.
The following steps require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
5. Return the slide-in module:
a) To request a dedicated packaging for the slide-in module and for
the appropriate procedure for returning the module, contact
your local Thermo Fisher Scientific support organization.
b) Pack the slide-in module in the dedicated packaging.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...