7
•
Maintenance and Service
Page 176
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Figure 77: Route and secure the needle capillary on the front surface of
the device
NOTICE
Do NOT over-tighten the attachment screw because the needle
capillary under the press plate may be damaged.
10. Install back the front plate with reverse steps. See
page 155).
11. Power on the device. Close the compartment doors and the device
will run initialization automatically. Make sure no errors or warnings
are reported in the Audit trail.
12. In Chromeleon, configure the
Needle Capillary
. There are two
options of the needle capillary types: 0.18 x 415 mm and 0.25 x 415
mm. Users can check via the marker on the needle capillary and
select accordingly. For more details, see
page 94)
.
13. In Chromeleon, select
Internal Needle Rinse
on ePanel to rinse the
internal surface of the needle capillary and needle unit with mobile
phase. For more details, see
Important Operating Parameters
(
page 113)
.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...