7
•
Maintenance and Service
Page 172
Integral Fraction Collector FT (VF-F20-A) Operating Manual
7.9
Replacing the Needle Capillary
When
If the needle capillary is clogged or damaged, or a new needle capillary
needs to be installed due to flow rate change.
Parts required
Replacement needle capillary
Figure 71: Needle capillary
No.
Description
No.
Description
1
Needle capillary Viper fitting
on the needle unit
5
S-shape slot on the foam part
2
Needle capillary screw
6
Needle capillary press plate
3
Needle capillary support block
7
Needle capillary Viper fitting on
the fraction collection valve
4
Needle capillary
Tools required
Screwdriver, Torx T10
Preparations
1. Prepare the device for maintenance, move the needle unit into
service position. See
page 153).
2. Turn off the device with its main power switch.