5
•
Installation
Page 94
Integral Fraction Collector FT (VF-F20-A) Operating Manual
5.11
Setting Up the Device in the Software
This manual assumes that the chromatography software is already
installed on the data system computer and a valid license is available.
For more information about setting up the
Vanquish system
in the
software, refer to the
Vanquish System Operating Manual
.
The Help for the software that you are using provides detailed
information about the settings on each property page.
Check the setting of the needle capillary
The
NeedleCapillary
property must be set correctly in the software
according to needle capillary installed on device, see
page 71)
. The needle capillary included in the
default shipment is ID 0.18 mm. This is also the default setting for
needle capillary in the Chromeleon.
If the needle capillary with ID 0.25 mm is installed the new capillary
dimensions must be set in Chromeleon accordingly.
Follow these steps
1. Check the marker physically located on the needle capillary. It is
either ID 0.18 mm or ID 0.25 mm. The default needle capillary
installed in the device is ID 0.18 mm.
2. In Chromeleon Console click
Command
icon or press F8.
Figure 38: Selecting the Command configuration page
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...