5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 91
2. Connect the delay capillary between the detector flow cell outlet or
the outlet (OUT) on the flow cell connection unit and the fraction
collection valve central port.
Figure 37: Connecting delay capillary to central port of fraction collection
valve
3. Route the delay capillary through the recess in the partition panel of
the detector.
4. Route the delay capillary between detector and the fraction
collector depending on length of the delay capillary that is used.
5.9.4
Connecting the Flush Buffer Loop
Parts required
Needle capillary
TIP
Make sure to check the flow rate of application and choose the
proper needle capillary according to application flow rate range.
Meanwhile, the installed flush buffer loop must be compatible with
specific needle capillary. See below table for details.
Flow rate
Needle Capillary
ID
Flush Buffer Loop
Volume
Remark
0.05
–
5 mL/min
0.18 mm
50 µL
Included in ship kit
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...