7
•
Maintenance and Service
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 155
7.6
Removing the Front Plate
When
1. For needle capillary installation
2. For easier access to inner fluidic components
3. For easier access to needle unit
4. For replacement if case of broken front plate cable
Parts required in case repair
Replacement front plate
Tools required
Screwdriver, Torx T10
Preparations
Turn off the device with its main power switch.
NOTICE
The device shall be powered off while replacing the front plate.
Follow these steps
1. Loosen the two screws that attaches the removable front plate with
the screwdriver. The screws shall remain in the front plate.
Figure 57: Loosening the two screws in front plate