Device Overview
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 35
switched into the flow path between detector and needle allowing to
dispense remaining fraction in the needle capillary and needle into the
current
tube
to increase the recovery of the fraction and reduced carry
over of the first fraction into the second fraction. When flush function is
completed the fraction collection valve will be switched to waste
position until a new fraction collection is triggered.
To remove residual sample from the needle external between runs, the
needle moves to the wash port and descends into it. The needle wash
pump supplies needle wash liquid to the wash port to wash the outer
surface of the needle to the wash waste.
To remove residual fraction from needle internal between runs, the
needle moves to the wash port and descends into it with fraction
collection valve located at collection position. The HPLC pump flow
supplies clean mobile phase to rinse the inner surface of the needle to
the wash waste.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...