6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 103
LED Bar
Description
Yellow
The device is connected in the Chromeleon software but is not
equilibrated.
Green, flashing
The device is equilibrating.
If you use fraction collector compartment thermostatting, the
thermostatting temperature is not yet achieved.
Green
The device is equilibrated, but no data acquisition is running.
If you use fraction collector compartment thermostatting, the
thermostatting temperature is achieved.
Blue, running
The device performs fraction collecting while a sequence is
running.
Blue
A sample or sequence is running, including data acquisition.
Red
A problem or error has occurred. For the related message, check
the Chromeleon Audit Trail. For remedial action, see the
Troubleshooting section in this operating manual.
STATUS LED
The
STATUS
LED on the left side of keypad provides the following
information:
STATUS LED
Description
Off (dark)
The power to the device is turned off.
Green
The device is functioning properly.
Red
A problem or error has occurred. For the related message, check
the Chromeleon Audit Trail. For remedial action, see
page 202)
.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...