Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 45
automatically verifies the rack type and orientation. The information is
sent to the Chromeleon software.
To allow the identification of the sample rack or well plate type, use
sample racks and well plates with such barcodes for rack type
identification.
Each segment accommodates a barcode label that informs the barcode
reader that the segment is empty if no sample rack or well plate is
installed. See
Carousel (
page 44)
.
Figure 12: Sample rack with barcodes for rack type identification
No.
Description
1
Position A1
2
Vanquish rack type barcode
3.4.4
Condensation Drainage
A leak sensor is embedded inside the fraction collector compartment.
The sensor detects liquid accumulated inside the fraction collector
compartment due condensing water when the device is operated in
cooling mode under humid conditions.
If liquid is detected, the drain pump is triggered immediately to transfer
the liquid to the drain port, from which the liquid is guided to the waste
reservoir through the drain system of the Vanquish system. If the liquid
is drained and the sensor gets dry, then the drain pump will resume idle
state. For more details, see
Messages (Code 9004) (
page 205)
.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...