7
•
Maintenance and Service
Page 180
Integral Fraction Collector FT (VF-F20-A) Operating Manual
4. Check if the needle unit is positioned correctly:
Move the lock pin slightly upward. If the lock pin and the vial pusher
can move up and down easily, the needle unit sits correctly in place.
5. Reinstall the needle capillary. See
Replacing the Needle Capillary
(
page 172)
.
6. Install the front plate. See
page 155)
.
7. Power on the device. Close the compartment doors and the device
will run initialization automatically. Make sure no errors or warnings
are reported in the Audit trail.