5
•
Installation
Page 86
Integral Fraction Collector FT (VF-F20-A) Operating Manual
5.9
Flush System
5.9.1
Choosing the Flush Solvent
Use the flush solvent that fulfills the following requirements:
•
Before filling the flush solvent reservoir, rinse the reservoir
thoroughly. Make sure that no particles, dust or algae are present.
•
Recommended when using 100% water as flush solvent
Replace the flush solvent daily.
•
Use a flush solvent that is suitable for your application and that
removes residual fraction from the flush buffer loop sufficiently.
5.9.2
Connecting Flush Solvent Reservoir
Parts required
•
Flush solvent reservoir
•
Flush solvent tubing
•
Retaining guide, 1/16’’
•
Solvent line filter
•
Filter frit
•
Solvent line adaptor (
1/8’’
-
1/16’’
)
•
Reservoir cap
•
Cap plugs
Preparations
1. Prepare the flush solvent and flush solvent reservoir. Observe the
flush solvent guidelines in
page 86)
.
2. Assemble the filter holder and solvent line adapter, wearing
appropriate clean gloves.
a) Place the frit in the filter holder (bottom part).
b) Make sure that the frit is in a level position.
c) Screw the filter top to the filter bottom.
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...