Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
250
Table 12-7:
DC Power Supply Acceptable Levels
POWER
SUPPLY
VOLTAGE
CHECK
RELAY BOARD TEST POINTS
MIN V
MAX V
FROM
Test Point
TO
Test Point
NAME
#
NAME
#
PS1
+5
DGND
1
+5
2
+4.85
+5.25
PS1
+15
AGND
3
+15
4
+13.5
+16.0
PS1
-15
AGND
3
-15V
5
-13.5
-16.0
PS1
AGND
AGND
3
DGND
1
-0.05
+0.05
PS1
Chassis
DGND
1
Chassis
N/A
-0.05
+0.05
PS2
+12
+12V Ret
6
+12V
7
+11.8
+12.5
PS2
DGND
+12V Ret
6
DGND
1
-0.05
+0.05
12.7.3. I
2
C BUS
Operation of the I
2
C bus can be verified by observing the behavior of D1 on the relay
PCA & D2 on the Valve Driver PCA. Assuming that the DC power supplies are
operating properly, the I
2
C bus is operating properly if D1 on the relay PCA and D2 of
the Valve Driver PCA are flashing
There is a problem with the I
2
C bus if both D1 on the relay PCA and D2 of the Valve
Driver PCA are ON/OFF constantly.
12.7.4. LCD/DISPLAY MODULE
TOUCHSCREEN
INTERFACE
Assuming that there are no wiring problems and that the DC power supplies are
operating properly, the display screen should light and show the splash screen and other
indications of its state as the CPU goes through its initialization process.
12.7.5. RELAY PCA
The Relay PCA can be most easily checked by observing the condition of the status
LEDs on the Relay PCA (see Section 12.3.2), and using the
SIGNAL I/O
submenu
under the
DIAG
menu (see Section 12.1.3) to toggle each LED
ON
or
OFF.
If D1 on the Relay PCA is flashing and the status indicator for the output in question
(Heater power, Valve Drive, etc.) toggles properly using the Signal I/O function, then
the associated control device on the Relay PCA is bad.
Several of the control devices are in sockets and can be easily replaced. The following
table lists the control device associated with a particular function:
Table 12-8:
Relay PCA Control Devices
FUNCTION
CONTROL DEVICE
SOCKETED
All valves
U5
Yes
Reaction Cell Heater
K1
Yes
NO
2
NO Converter heater
K2
Yes
Permeation Tube Heater for
Optional Internal Span Gas Generator
K4
Yes
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...