APPENDIX A-3: Warnings and Test Measurements
Teledyne API - T200, T204 and 200E Series (05295F DCN6900)
A-14
Test Name
1
Message Text
Description
CO2CELLTEMP
8
CO2 CELL TEMP=50.8 C
CO
2
sensor cell temperature.
O2CELLTEMP
5
O2 CELL TEMP=50.8 C
O
2
sensor cell temperature.
IZSTEMP
IZS TEMP=50.8 C
IZS temperature.
CONVTEMP
MOLY TEMP=315.0 C
Converter temperature. Converter type
is MOLY, CONV, or O3KL.
SAMPRESTTEMP
10
SMP RST TMP=49.8 C
Sample restrictor temperature.
RCELLPRESS
RCEL=7.0 IN-HG-A
Reaction cell pressure.
SAMPPRESS
SAMP=29.9 IN-HG-A
Sample pressure.
NOXSLOPE
NOX SLOPE=1.000
NOX slope for current range, computed
during zero/span calibration.
NOXOFFSET
NOX OFFS=0.0 MV
NOX offset for current range,
computed during zero/span calibration.
NOSLOPE
NO SLOPE=1.000
NO slope for current range, computed
during zero/span calibration.
NOOFFSET
NO OFFS=0.0 MV
NO offset for current range, computed
during zero/span calibration.
NO2 NO2=0.0
PPB
3
NO
2
concentration for current range.
NO2_1
7
NO2_1=0.0
PPB
3
NO
2
concentration for range #1.
NO2_2
7
NO2_2=0.0
PPB
3
NO
2
concentration for range #2.
NOX NOX=396.5
PPB
3
NOX concentration for current range.
NOX_1
7
NOX_1=396.5
PPB
3
NOX concentration for range #1.
NOX_2
7
NOX_2=396.5
PPB
3
NOX concentration for range #2.
NO NO=396.5
PPB
3
NO concentration for current range.
NO_1
7
NO_1=396.5
PPB
3
NO concentration for range #1.
NO_2
7
NO_2=396.5
PPB
3
NO concentration for range #2.
CO2RANGE
8, not 6
CO2 RANGE=100.00 PCT
D/A #4 range for CO
2
concentration.
CO2SLOPE
8
CO2
SLOPE=1.000
CO
2
slope, computed during zero/span
calibration.
CO2OFFSET
8
CO2
OFFSET=0.000
CO
2
offset, computed during zero/span
calibration.
CO2
8
CO2=15.0
%
CO
2
concentration.
O2RANGE
5, not 6
O2 RANGE=100.00 PCT
D/A #4 range for O
2
concentration.
O2SLOPE
5
O2
SLOPE=1.000
O
2
slope computed during zero/span
calibration.
O2OFFSET
5
O2 OFFSET=0.00 %
O
2
offset computed during zero/span
calibration.
O2
5
O2=0.00
%
O
2
concentration.
TESTCHAN
5,6,8
TEST=3627.1 MV
Value output to TEST_OUTPUT analog
output, selected with TEST_CHAN_ID
variable.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...