Calibration Procedures
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
202
EXIT
returns
to the
SETUP
Menu.
SETUP
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Toggle buttons to set
duration for each iteration
of the sequence:
Set in Decimal minutes
from 0.1 – 60.0.
SETUP
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP
CALIBRATE: OFF
ON
ENTR
EXIT
SETUP X.X
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle button
Between Off and
ON.
SETUP X.X
SEQ 2) ZERO
–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Display show:
SEQ 2) ZERO
–SPAN, 2:00:30
Sequence
Delta Time
MODE Delta Days
CONTINUED FROM PREVIOUS PAGE
DELTA TIME
Note
If at any time an unallowable entry is selected (Example: Delta Days > 367) the
ENTR button will disappear from the display.
06858E DCN7057