Principles of Operation
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
324
13.7.1.1. AC CONFIGURATION
– INTERNAL PUMP (JP7)
If your T200 includes an internal pump the following table, jumper set JP7 is used to
configure the power supplied to it as shown in Figure 13-24.
Table 13-6: AC Power Configuration for Internal Pumps (JP7)
LINE
POWER
LINE
FREQUENCY
JUMPER
COLOR
FUNCTION
JUMPER
BETWEEN
PINS
110VAC
115 VAC
60 HZ
WHITE
Connects pump pin 3 to 110 / 115 VAC power line
2 to 7
Connects pump pin 3 to 110 / 115 VAC power line
3 to 8
Connects pump pins 2 & 4 to Neutral
4 to 9
50 HZ
1
BLACK
Connects pump pin 3 to 110 / 115 VAC power line
2 to 7
Connects pump pin 3 to 110 / 115 VAC power line
3 to 8
Connects pump pins 2 & 4 to Neutral
4 to 9
220VAC
240 VAC
60 HZ
BROWN
Connects pump pins 3 and 4 together
1 to 6
Connects pump pin 1 to 220 / 240VAC power line
3 to 8
50 HZ
1
BLUE
Connects pump pins 3 and 4 together
1 to 6
Connects pump pin 1 to 220 / 240VAC power line
3 to 8
1
A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the Model T200.
110 VAC /115 VAC
220 VAC /240 VAC
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
May be present on 50 Hz version of
jumper set, but is not functional on
the T200
Figure 13-24: Pump AC Power Jumpers (JP7)
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...