Setup Mode Menus
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
110
5.9.3.5. AUTOMATIC INDIVIDUAL CALIBRATION OF THE ANALOG OUTPUTS
Use the
AUTO CAL
feature to initiate an automatic calibration for an individual analog
output;
access
the
ANALOG
I/O
CONFIGURATION
submenu
(SETUP>MORE>DIAG>PASSWORD>NEXT . . . or see Figure 5-5) then press:
From the
AIO CONFIGURATION SUBMENU
DIAG AIO CONC_OUT_2: 5V, CONC2, NOCAL
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
DIAG AIO CONC_OUT_2: CALIBRATED:NO
<SET
SET>
CAL
EXIT
DIAG AIO CONC_OUT_2: RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
DIAG AIO CONC_OUT_2: CALIBRATED: YES
<SET
SET> CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
06858E DCN7057