Calibration Procedures
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
188
TO turn the
CAL_ON_NO
2
feature
ON/OFF
, press:
Use this button to
turn this feature
ON/OFF.
Continue pressing
NEXT
until ...
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST>
CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
<PREV
NEXT>
JUMP
EDIT PRNT EXIT
SETUP X.X
09)CAL_ON_NO2=OFF
<PREV NEXT> JUMP
EDIT
PRNT EXIT
MODE FLD
CAL_ON_NO2=OFF
CONC
ON
ENTR
EXIT
Press
EXIT
3 times.
06858E DCN7057