Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Setup Mode Menus
97
5.6.2. ADJUSTING THE I
NTERNAL CLOCK’S SPEED
In order to compensate for CPU clocks that run fast or slow, adjust a variable called
CLOCK_ADJ
to speed up or slow down the clock by a fixed amount every day.
The
CLOCK_ADJ
variable is accessed via the
VARS
submenu: To change the value of
this variable, press:
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT ENTR EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X ENTER SETUP PASS:0
8
1
8
ENTR
EXIT
Continue pressing
NEXT
until ...
SETUP X.X 8) CLOCK_ADJUST=0 Sec/Day
+
0
0
EDIT
ENTR
EXIT
SETUP X.X 8) CLOCK_ADJUST=0 Sec/Day
PREV NEXT
EDIT
ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+)
SETUP X.X 8) CLOCK_ADJUST=0 Sec/Day
PREV NEXT JUMP
EDIT
ENTR EXIT
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
Concentration field
displays all gases.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST> CAL
SETUP
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...