Setup Mode Menus
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
84
A2
OUTPUT: NO Output Signal = 0 – 10 VDC representing 0-500 ppb concentration
values.
A3
OUTPUT: NO
2
Output Signal = 0 – 5 VDC representing 0-500 ppb concentration
values.
The user may also add a signal offset independently to each output (see Section 5.9.3.9)
to match the electronic input requirements of the recorder or data logger to which the
output is connected.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
The instrument does not remember upper range limits settings associated with
the individual modes. Changes made to the range limits (e.g. 400 ppb
600 ppb)
when in one particular mode will alter the range limit settings for the other
modes.
When switching between reporting range modes, ALWAYS check and reset the
upper range limits for the new mode selection..
5.4.2.2. ANALOG OUTPUT REPORTING RANGE DEFAULT SETTINGS
The default setting for these the reporting ranges of the analog output channels
A1, A2
and
A3
are:
SNGL mode
0 to 500.0 ppb
0 to 5 VDC
5.4.3. SETUP
RNGE
MODE
Single range mode (SNGL) reports all three of the NOx gas concentrations using the
same reporting range span (see Section 5.4.3.1).
Independent range mode (IND) allows the NOx, NO and NO
2
analog outputs to be set
with different reporting range spans (see Section 5.4.3.2).
Automatic range mode (AUTO) allows the analyzer to automatically switch the
reporting range between two user upper span limits (designated LOW and HIGH) based
on the actual concentrations being measured for each (see Section 5.4.3.3). These are not
the same as the analyzer‟s low and high physical ranges (Section 5.4.1).
5.4.3.1. SETUP
RNGE
MODE
SNGL: CONFIGURING THE T200 ANALYZER FOR SINGLE
RANGE MODE
Note
Single Range is the default reporting range mode for the analyzer.
When the single range mode is selected (
SNGL
), all analog NO
x
, NO and NO
2
concentration outputs (
A1, A2 and A3
) are slaved together and set to the same reporting
range limits (e.g. 500.0 ppb). This reporting range can be set to any value between 100
ppb and 20,000 ppb.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...