Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Getting Started
53
P
NEUMATIC
L
AYOUT FOR
B
ASIC
C
ONFIGURATION
SAMPLE
PRESSURE
SENSOR
VACUUM
PRESSURE
SENSOR
O
3
FLOW
SENSOR
FLOW PRESSURE
SENSOR PCA
INSTRUMENT CHASSIS
PMT
NO/NO
X
VALVE
AUTOZERO
VALVE
O
3
Destruct
O
3
GENERATOR
O
3
Cleanser
EXHAUST
GAS OUTLET
SAMPLE
GAS INLET
NO
2
Converter
Orifice Dia.
0.004"
Orifice Dia.
0.010"
PUMP
NO
X
Ex
h
a
u
s
t
Sc
ru
b
b
e
r
OZONE DRYER
F
ilt
e
r
Orifice Dia.
0.004"
EXH
AU
ST
M
AN
IF
O
L
D
Orifice Dia.
0.010"
COM
NC
NO
NC
COM
NO
Figure 3-19:
Pneumatics, Basic Configuration
3.3.2.3. CONNECTIONS W/AMBIENT ZERO/AMBIENT SPAN (Z/S) VALVES (OPT 50A)
This valve package includes:
two solenoid valves located inside the analyzer that allow the user to switch either
zero, span or sample gas to the instrument’s sensor
two additional gas inlet ports (
ZERO AIR
and
SPAN1
)
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...