Teledyne API - T200, T204 and 200E Series (05295F DCN6900)
Appendix A3: Warnings and Test Measurements, Software Version K.3
A-23
Table A-6: DAS Parameters
Name
Description
Units
PMTDET
PMT detector reading
mV
RAWNOX
6
Raw PMT detector reading for NOX
mV
RAWNO
6
Raw PMT detector reading for NO
mV
NXSLP1 NO
X
slope for range #1
—
NXSLP2 NO
X
slope for range #2
—
NXSLP3
7
NO
X
slope for range #3
—
NOSLP1
NO slope for range #1
—
NOSLP2
NO slope for range #2
—
NOSLP3
7
NO slope for range #3
—
NXOFS1 NO
X
offset for range #1
mV
NXOFS2 NO
X
offset for range #2
mV
NXOFS3
7
NO
X
offset for range #3
mV
NOOFS1
NO offset for range #1
mV
NOOFS2
NO offset for range #2
mV
NOOFS3
7
NO offset for range #3
mV
CO2SLP
5
CO
2
slope
—
CO2OFS
5
CO
2
offset
%
O2SLPE
3
O
2
slope
—
O2OFST
3
O
2
offset
%
NXZSC1 NO
X
concentration for range #1 during zero/span
calibration, just before computing new slope and offset
PPB
2
NXZSC2 NO
X
concentration for range #2 during zero/span
calibration, just before computing new slope and offset
PPB
2
NXZSC3
7
NO
X
concentration for range #3 during zero/span
calibration, just before computing new slope and offset
PPB
2
NOZSC1
NO concentration for range #1 during zero/span calibration,
just before computing new slope and offset
PPB
2
NOZSC2
NO concentration for range #2 during zero/span calibration,
just before computing new slope and offset
PPB
2
NOZSC3
7
NO concentration for range #3 during zero/span calibration,
just before computing new slope and offset
PPB
2
N2ZSC1 NO
2
concentration for range #1 during zero/span
calibration, just before computing new slope and offset
PPB
2
N2ZSC2 NO
2
concentration for range #2 during zero/span
calibration, just before computing new slope and offset
PPB
2
N2ZSC3
7
NO
2
concentration for range #3 during zero/span
calibration, just before computing new slope and offset
PPB
2
CO2ZSC
5
CO
2
concentration during zero/span calibration, just before
computing new slope and offset
%
O2ZSCN
3
O
2
concentration during zero/span calibration, just before
computing new slope and offset
%
NXCNC1 NO
X
concentration for range #1
PPB
2
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...