Setup Mode Menus
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
90
To set individual ranges press the following menu sequence:
Concentration field
displays all gases.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
RANGE MODE MENU
MODE
SET UNIT DIL
EXIT
SETUP X.X
RANGE MODE:SNGL
SNGL IND
AUTO
DIL
ENTR EXIT
SETUP X.X
RANGE MODE:AUTO
SNGL IND AUTO
ENTR
EXIT
SETUP X.X
RANGE MODE MENU
MODE
SET
UNIT DIL
EXIT
SETUP X.X
HIGH RANGE:200.0 Conc
0
0
5
0
0
.0
ENTR EXIT
Toggle these buttons
to select the upper
SPAN
limit for the
reporting range.
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
LOW RANGE:50.0 Conc
0
0
5
0
0
.0
ENTR EXIT
The
LOW
and
HIGH
ranges have separate
slopes and offsets for
computing the NO
X
and NO concentration.
The two ranges must
be independently
calibrated
.
NOTE:
Avoid accidentally setting the
LOW
range (
RANGE1
) of the instrument
with a higher span limit than the
HIGH
range (
RANGE2
).
This will cause the unit to stay in
the low reporting range perpetually
and defeat the function of the
AUTO
range mode.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...