Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
260
Converter Efficiency Menu
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST>
CAL
SETUP
SAMPLE
RANGE TO CAL
LOW HIGH
ENTR
EXIT
M-P CAL
CONCENTRATON MENU
NOX
NO
CONV
EXIT
M-P CAL
CONVERTER EFICIENCY MENU
NO2
CAL
SET
EXIT
M-P CAL
NO2 CE CONC: 500.0 Conc
0
4
0
0.
0
0
ENTR EXIT
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST>
CAL
SETUP
This step only appears if the
analyzer’s reporting range is
set for
AUTO
range mode.
Select
LOW
and press
ENTR
.
Repeat entire procedure for
HIGH
range.
The expected NO
2
span
concentration value defaults
to
400.0 Conc
.
Make sure that you specify
the actual concentration value
of the NO
2
gas.
Toggle these buttons
to change this value to
the concentration of
the NO
2
gas being
used.
Use these buttons to
select the appropriate
range.
Repeat entire procedure
for each range.
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST> ZERO
CONC
SETUP
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...