13.7.1.2. AC CONFIGURATION
– STANDARD HEATERS (JP2)
Power configuration for the AC the standard heaters is set using Jumper set JP2 (see
Figure 13-25 for the location of JP2).
Table 13-7: Power Configuration for Standard AC Heaters (JP2)
LINE VOLTAGE
JUMPER
COLOR
HEATER(S)
JUMPER
BETWEEN
PINS
FUNCTION
110 VAC / 115 VAC
50Hz & 60 Hz
WHITE
Reaction Cell / Sample
Chamber Heaters
1 to 8
Common
2 to 7
Neutral to Load
4 to 9
Neutral to Load
Moly Converter
3 to 10
Common
4 to 9
Neutral to Load
6 to 11
Neutral to Load
220 VAC / 240 VAC
50Hz & 60 Hz
BLUE
Reaction Cell / Sample
Chamber Heaters
1 to 7
Load
Moly Converter
3 to 9
Load
1
2
3
4
5
6
7
8
9
10
11
12
Reaction Cell
Heaters
NO
2
NO
Converter Heaters
110 VAC /115 VAC
1
2
3
4
5
6
7
8
9
10
11
12
220 VAC / 240 VAC
Reaction Cell
Heaters
NO
2
NO
Converter Heaters
Figure 13-25: Typical Set Up of AC Heater Jumper Set (JP2)
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...