Analyzer Operation Manual
Calibration Procedures
191
To set the span gas concentration, press:
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
M-P CAL
RANGE=500.0 PPB NOX= XXXX
<TST TST> ZERO SPAN
CONC
EXIT
M-P CAL CONCENTRATION MENU
NOX
NO
CONV
EXIT
M-P CAL NOX SPAN CONC:80.0 Conc
0
4
0
0
.0
ENTR EXIT
Enter the
concentration
measured during
the last span check
of the permeation
tube output.
EXIT
ignores the new setting and
returns to the previous display.
ENTR
accepts the new setting and
returns to the
CONCENTRATION MENU.
If using NO span gas in addition to NO
X
repeat last step.
Only appears if the
AUTO
range mode is selected.
Use these buttons to select
the appropriate range.
Repeat entire procedure for
each range.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST> CAL CALZ
CALS
SETUP
06858E DCN7057