Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Getting Started
67
3.4.3. FUNCTIONAL CHECKS
After the analyzer‟s components have warmed up for at least 60 minutes, verify that the
software properly supports any hardware options that are installed.
For information on navigating through the analyzer‟s software menus, see the menu
trees illustrated in Appendix A.1.
Check to ensure that the analyzer is functioning within allowable operating parameters.
Appendix C includes a list of test functions viewable from the analyzer’s front panel
as well as their expected values.
These functions are also useful tools for diagnosing problems with your analyzer.
The enclosed Final Test and Validation Data sheet (P/N 04409) lists these values
before the instrument left the factory.
To view the current values of these parameters press the following button sequence on
the analyzer‟s front panel. Remember until the unit has completed its warm up these
parameters may not have stabilized.
1
This will match the currently selected units of measure
for the range being displayed.
2
The
STB
function can be set to display data related to
any of the gases the analyzer measures, e.g. NO
X
,
NO, NO
2
or O
2
(if the O
2
sensor option is installed.
3
Only appears if IZS option is installed.
4
Only appears if analog output
A4
is actively reporting
a
TEST FUNCTION
SAMPLE
RANGE=500.0 PPB NO= XXXX
<TST
TST> CAL
SETUP
Toggle
<TST TST>
buttons
to scroll through list of
functions
RANGE=
[Value]
PPB
1
RANGE1=
[Value]
PPB
1
RANGE2=
[Value]
PPB
1
NOX STB=
[Value]
PPB
2
SAMP FLW=
[Value]
CC/M
OZONE FL=
[Value]
CC/M
PMT=
[Value]
MV
NORM PMT=
[Value]
MV
AZERO=
[Value]
MV
HVPS=
[Value]
V
RCELL TEMP=
[Value]
ºC
BOX TEMP=
[Value]
ºC
PMT TEMP=
[Value]
ºC
IZS TEMP=
[Value]
ºC
3
MOLY TEMP=
[Value]
ºC
RCEL=
[Value]
IN-HG-A
SAMP=
[Value]
IN-HG-A
NOX SLOPE=
[Value]
NOX OFFS=
[Value]
MV
NO SLOPE=
[Value]
NO OFFS=
[Value]
MV
TEST=
[Value]
MV
4
TIME=
[HH:MM:SS]
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...