Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Setup Mode Menus
95
SYSTEM ENTER SETUP PASS:0
8
1
8
ENTR
EXIT
SYSTEM ENTER SETUP PASS:0
0
0
0
ENTR EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
EXAMPLE: This
password enables the
SETUP
mode.
Press individual
buttons to set
number.
Instrument enters selected menu.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
Note
The instrument still prompts for a password when entering the VARS and DIAG
menus, even if passwords are disabled, but it displays the default password (818)
upon entering these menus. In this case, the user only has to press ENTR to
access the password-protected menus.
In order to disable the PASSWORD feature after it has been turned ON, go to the
SETUP menu, input the password, and press ENTR. In the Primary Setup Menu press
PASS; then press ON to turn PASSWORD ENABLE back to OFF, and press ENTR to
accept and apply the change (no need to press EXIT from the Primary Setup Menu this
time).
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...