Setup Mode Menus
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
116
5.9.3.8. TURNING AN ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF
In its default configuration, a ± 5% over-range is available on each of the T200‟s analog
outputs. This over-range can be disabled if your recording device is sensitive to excess
voltage or current.
To turn the over-range feature on or off, select the
ANALOG I/O CONFIGURATION
submenu (see Figure 5-5) then press:
DIAG AIO CONC_OUT_2: 5V, OVR, NOCAL
<SET
SET>
EDIT
EXIT
DIAG AIO CONC_OUT_2: OVERRANGE: ON
<SET
SET>
EDIT
EXIT
DIAG AIO CONC_OUT_2: RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO CONC_OUT_2: OVERRANGE: ON
ON
ENTR EXIT
DIAG AIO CONC_OUT_2: OVERRANGE: OFF
OFF
ENTR
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Toggle this
button to turn
the Over-
Range feature
ON/OFF.
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
From the
AIO CONFIGURATION SUBMENU
06858E DCN7057