Attach an exhaust line to the EXHAUST outlet fitting. The exhaust line should be:
¼” PTFE tubing
a maximum of 10 meters long
vented outside the T200 analyzer’s enclosure
P
NEUMATIC
L
AYOUT FOR
A
MBIENT
Z
ERO
/P
RESSURIZED
S
PAN
(OPT
50B)
SAMPLE
PRESSURE
SENSOR
VACUUM
PRESSURE
SENSOR
O
3
FLOW
SENSOR
FLOW PRESSURE
SENSOR PCA
INSTRUMENT CHASSIS
PMT
OZONE DRYER
NO/NO
X
VALVE
AUTOZERO
VALVE
O
3
Destruct
O
3
GENERATOR
O
3
Cleanser
Filter
NO
2
Converter
ZERO GAS
INLET
PRESSURIZED
SPAN GAS INLET
SAMPLE/ CAL
VALVE
ZERO/SPAN
VALVE
2
-S
ta
ge
NO
X
S
cr
ub
be
r
Exhaust
Outlet
NO
NC
NO
NC
NO
NC
C
O
M
NO
NC
C
O
M
NC
BYPASS SPAN
GAS OUTLET
SAMPLE
GAS
INLET
PRESSURIZED SPAN
OPTION
SPAN GAS
DO NOT EXCEED 30 PSIG
EXH
AU
ST
M
AN
IF
O
L
D
Orifice Dia.
0.013"
Orifice Dia.
0.010"
Orifice Dia.
0.010"
Orifice Dia.
0.004"
PUMP
NO
X
Ex
h
a
u
s
t
Sc
ru
b
b
e
r
Orifice Dia.
0.004"
COM
NO
COM
COM
Figure 3-25:
Pneumatics with Ambient Zero/Pressurized Span Valves (OPT 50B)
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...