Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
236
12.3. USING THE INTERNAL ELECTRONIC STATUS LEDS
Several LEDs are located inside the instrument to assist in determining if the analyzer‟s
CPU, I
2
C bus and Relay PCA are functioning properly.
12.3.1. CPU STATUS INDICATOR
DS5, a red LED, that is located on upper portion of the motherboard, just to the right of
the CPU board, flashes when the CPU is running the main program loop. After power-
up, approximately 30 – 60 seconds, DS5 should flash on and off. If characters are
written to the front panel display but DS5 does not flash then the program files have
become corrupted, contact Teledyne API's Customer Service Department (see Section
12.10) because it may be possible to recover operation of the analyzer. If after 30 – 60
seconds, neither DS5 is flashing nor have any characters been written to the front panel
display then the CPU is bad and must be replaced.
Motherboard
CPU Status LED
Figure 12-2:
CPU Status Indicator
12.3.2. RELAY PCA STATUS LEDS
There are sixteen LEDs located on the Relay PCA. Some are not used on this model.
12.3.2.1. I
2
C BUS WATCHDOG STATUS LEDS
The most important is D1 (see Figure 12-3), which indicates the health of the I
2
C bus.
Table 12-4: Relay PCA Watchdog LED Failure Indications
LED
Function
Fault Status
Indicated Failure(s)
D1
(Red)
I
2
C bus Health
(Watchdog Circuit)
Continuously ON
or
Continuously OFF
Failed/Halted CPU
Faulty Motherboard, Touchscreen or Relay PCA
Faulty Connectors/Wiring between Motherboard,
Touchscreen or Relay PCA
Failed/5 VDC Power Supply (PS1)
If D1 is blinking, then the other LEDs can be used in conjunction with
DIAG
Menu
Signal I/O to identify hardware failures of the relays and switches on the Relay PCA.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...