Specifications
4- 12
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP404XL Maximum Measurement Times
Figures 4--8 through 4--10 show the maximum measurement time for the
TCP404XL probe, with varying duty cycles and temperatures. For more detail
on the relationship between measurement factors, refer to
Measuring Noncontin-
uous Current with the TCP404XL Probe
on page 3--6.
Measuring 750A at 50
_
C ambient
Duty cycle (%)
0
5
10
15
20
25
30
35
0
2
4
6
8
10
12
M
axim
um
m
easurem
ent
time
(m
in)
0A Continuous
100A Continuous
200A Continuous
300A Continuous
Figure 4- 8: Measuring 750A noncontinuous at 50
°
C ambient temperature
Duty cycle (%)
0
10
20
30
40
50
60
0
5
10
15
20
25
30
M
axim
um
m
easurem
ent
time
(m
in)
Measuring 600A at 50
_
C ambient
70
0A Continuous
300A Continuous
400A Continuous
200A Continuous
Figure 4- 9: Measuring 600A noncontinuous at 50
°
C ambient temperature
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......