TCP305 and TCP312 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 21
DC Gain Accuracy
This procedure tests the DC gain accuracy of the TCP305 and TCP312 Current
Probes. In this test you compare the voltage output of the amplifier to a reference
input.
This test uses current loops that you make using wire and 3-inch forms. Refer to
page 5--4 for instructions on making the current loops.
1.
Using a BNC cable, connect the amplifier
OUTPUT
to the 50
Ω
feed-
through termination. Attach the termination to a BNC-to-dual banana
adapter. Insert the dual-banana adapter into the digital multimeter DC
voltage input, taking care to maintain polarity.
2.
Connect the current loop to the current source as shown in Figure 5--7.
3.
Connect the current probe to the
PROBE INPUT
connector. Do not connect
the probe to the current loop at this time.
Current
source
DMM
+ --
+
--
5 Turns
Probe
BNC-to-dual
banana adapter
50
Ω
Precision
termination
Output
50
Ω
Coaxial
cable
(Degauss before
connecting to current loop)
Probe
Current flow
Amplifier
Figure 5- 7: DC gain accuracy test setup for TCP305 and TCP312
Make or verify the equipment settings in Table 5--12:
Table 5- 12: Equipment settings for DC gain accuracy
Digital multimeter
Measurement Type
DC volts
Range
Autoranging
Current source
Amplitude
DC Amps
Equipment Connections
Equipment Settings
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......