Getting Started
1- 2
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
The amplifier amplifies the current sensed by the probe and converts the current
to a proportional voltage that is displayed on an oscilloscope or other similar
measuring device.
The following Tektronix current probes are compatible with the TCPA300
Amplifier:
H
TCP312 (30 amps, 100 MHz)
H
TCP305 (50 amps, 50 MHz)
H
TCP303 (150 amps, 15 MHz)
The following Tektronix current probes are compatible with the TCPA400
Amplifier:
H
TCP404XL (750 amps*, 2 MHz)
* 500 amps continuous, 750 amps DC derated with duty cycle
You can also use the CT-4 High-Current Transformer with the TCP305 and
TCP312 current probes to extend the AC current measurement range to
20,000 peak amps.
An oscilloscope displays the output from the current measuring system. A 50
Ω
cable is included to connect the amplifier to the oscilloscope input channel. A
TEKPROBE-to-TEKPROBE interface cable is also included for connecting to
TEKPROBE level II oscilloscopes.
If the oscilloscope does not have an input that can be set to 50
Ω
impedance, you
need a feedthrough 50
Ω
termination. This termination is included as a standard
accessory with your TCPA300 and TCPA400 Current Probe Amplifiers.
TCPA300 and TCPA400
Current Probe Amplifiers
Current Probes
Oscilloscope
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......