TCP305 and TCP312 Adjustments
6- 8
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP305 and TCP312 DC Gain Adjustment
Use the following procedure to adjust the DC Gain of the TCP305 and TCP312
Probes.
1.
Connect the amplifier
OUTPUT
to the DMM input using a BNC-to-Dual
Banana connector, 50
Ω
BNC cable, and a 50
Ω
precision feedthrough
termination. Do not connect the termination at the amplifier output.
2.
Connect the current probe to the amplifier
PROBE INPUT
connector.
3.
Connect the 5-turn DC current loop to the current source as shown in
Figure 6--4 on page 6--9.
4.
Do not clamp the current probe around any conductor, but make sure the
slide is locked.
Make or verify the equipment settings in Table 6--6:
Table 6- 6: Settings for DC gain adjustment
Digital multimeter
Measurement Type
AC volts
Range
Autoranging
Current source
Output waveshape/frequency
Square wave, 40 Hz
Output
Off
TCPA300
Coupling
DC
Range:
TCP305
5 A/V
TCP312
1 A/V
Equipment Connections
Equipment Settings
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......