TCP303 and TCP404XL Performance Verification
5- 42
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP303 Current Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP303 test record
Instrument Serial Number:
Certificate Number:
Temperature:
Relative Humidity %:
Date of Calibration:
Technician:
Performance test
Range, A/V
Test
current, A
Minimum
Measured/calculated
Maximum
DC Gain Accuracy
1
(% Error)
5 A/V
1
--3%
+3%
/
3
--3%
+3%
5
--3%
+3%
50 A/V
0.6
--3%
+3%
/
1.8
--3%
+3%
3.0
--3%
+3%
Rise Time
5 A/V
------
------
23 ns
Bandwidth
5 A/V
------
.707
------------
1
DC Gain Accuracy is tested at the 10
°
C to 50
°
C specification. If you need to test the probe at 0
°
C to 10
°
C, refer to the
specification limits on page 4- 1.