TCPA300 and TCPA400 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 7
3.
Make or verify the equipment settings in Table 5--4:
Table 5- 4: Equipment settings
Digital multimeter
Measurement Type
DC volts
Range
Autoranging
Current source
Voltage
6 V
Current
100 mA
Output
Off
TCPA300 and TCPA400 amplifier
Coupling
DC
After the equipment is set up, proceed as follows:
4.
Connect the Calibration Adapter to the
PROBE INPUT
of the amplifier.
5.
For each of the Range settings in Table 5--5, perform the following steps:
a.
If you are checking a TCPA300, set the amplifier to the desired Range
setting in Table 5--5.
b.
Enable the output of the current source.
c.
Record the exact measurement of the digital multimeter as
M
1
.
Table 5- 5: DC gain accuracy test for the TCPA300 and TCPA400
Amplifier
Range, A/V
Current
source out-
put
Expected
output, V
E
(VDC)
Measured
output, M
1
(VDC)
% Error,
calculated
TCPA300
1
100 mA
5.0000
5
100 mA
2.5000
10
100 mA
1.2375
50
100 mA
0.4835
50 (COMP)
1
100 mA
0.4585
TCPA400
1 A/mV
100 mA
2.5000
1
Both 10 A/V and 50 A/V RANGE LEDs light.
Procedure
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......