TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 19
TCP305 and TCP312 Performance Verification
The test procedures in this section verify warranted specifications and proper
performance of the TCP305 and TCP312 current probes, using the TCPA300
Amplifier. Tolerances that are specified in these procedures apply to the amplifier
and current probe and do not include test equipment error.
Before starting these procedures, photocopy the appropriate test record beginning
on page 5--29 to record the performance test results. The recommended
calibration interval is one year.
NOTE
. If the amplifier is not properly terminated into 50
Ω
., the NOT TERMI-
NATED INTO 50
Ω
status LED is on. Verify that the 50
Ω
termination is
connected at the DMM input when performing the following procedures.
No probe or amplifier adjustments are required during these test procedures. If
any tests fail, refer to
TCP305 and TCP312 Adjustments
, beginning on page 6--7.
WARNING.
Magnetic fields are produced in these procedures that may cause a
malfunction in heart pacemakers or damage to sensitive equipment.
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......