TCP305 and TCP312 Performance Verification
5- 30
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP312 Current Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP312 test record
Instrument Serial Number:
Certificate Number:
Temperature:
Relative Humidity %:
Date of Calibration:
Technician:
Performance test
Range, A/V
Test
current, A
Minimum
Measured/calculated
Maximum
DC Gain Accuracy (% Error)
1 A/V
0.20 A
--3%
+3%
/
0.60 A
--3%
+3%
1.00 A
--3%
+3%
10 A/V
1.20 A
--3%
+3%
/
3.60 A
--3%
+3%
6.00 A
--3%
+3%
Rise Time
1 A/V
------
≤
3.5 ns
Bandwidth
1 A/V
.707
------------