Control Summary
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
2- 3
Displaying Error Codes with the Probe Degauss Autobalance Button
on
page 3--22.
NOTE
. The NOT TERMINATED INTO 50
Ω
LED is on when the amplifier output
is not properly terminated into a 50
Ω
load. Make sure your amplifier OUTPUT
is connected to an oscilloscope input using a 50
Ω
BNC cable, and that the
oscilloscope input is set to 50
Ω
impedance. See Figure 1--1 on page 1--1 for
proper cabling.
If your oscilloscope does not have 50
Ω
impedance settings for inputs, you can
place a 50
Ω
feed-through termination on the oscilloscope input and connect the
amplifier output cable to the termination. Do not place the feed-through
termination at the amplifier end of the BNC connecting cable.
The MANUAL BALANCE buttons allow you to fine-adjust the DC offset that
appears at the amplifier OUTPUT connector. The manual balance adjustment
only functions when the amplifier is set to DC coupling, and the MANUAL
BALANCE indicator is only lighted after you press one of the MANUAL
BALANCE buttons in DC coupling mode.
When lit, this indicator informs you that the current probe is unlocked. You must
have the probe slide locked to degauss the probe or to accurately measure
current.
When this LED is red, it informs you that the measurement you are taking
exceeds the
continuous
amplitude limit of the TCPA300 and TCPA400. The
detection circuits detect only low frequency and DC overloads. Since overloads
can magnetize the probe, always degauss the probe after an overload. When the
OVERLOAD indicator flashes red, it indicates the pulsed waveform is exceeding
the amplitude limit of the amplifier.
When this LED is orange, it indicates that the safe operating temperature of the
probe, and possibly the amplifier, has been exceeded. Disconnect the probe from
the current source and allow time for the probe head and amplifier to cool.
When this LED blinks red and orange, it indicates that both the safe operating
temperature of the probe and the current limit have been exceeded.
WARNING.
To avoid personal injury or equipment damage, do not exceed the
specified electrical limits of the TCPA300 and TCPA400 or any applicable
accessories.
MANUAL BALANCE
Buttons and Indicator
PROBE OPEN Indicator
OVERLOAD Indicator
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......