Specifications
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 5
Environmental Characteristics
The environmental characteristics in Table 4--6 are warranted performance
specifications. Unlike the warranted characteristics in Table 4--1, the environ-
mental characteristics are type tested; therefore there are no performance
verification procedures provided to test these characteristics. Unless otherwise
noted, these characteristics apply to all probes and amplifiers.
Table 4- 6: Environmental characteristics
Parameter
Characteristic
Ambient Temperature
Operating
0
_
C to +50
_
C (32
_
F to 122
_
F)
Nonoperating
--40
_
C to +75
_
C (--40
_
F to 167
_
F)
Humidity
Operating
5 to 95% R.H. to +30
_
C (86
_
F)
5 to 85% R.H. +30
_
C to +50
_
C (86
_
F to 122
_
F)
Nonoperating
5 to 95% R.H. to +30
_
C (86
_
F)
5 to 85% R.H. +30
_
C to +75
_
C (86
_
F to 167
_
F)
Altitude
Operating
2000 m (6800 ft) maximum
Nonoperating
12,192 m (40,000 ft) maximum
Random Vibration, Amplifiers
Operating
0.31 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Nonoperating
2.46 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Random Vibration, Probes
Operating
2.66 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Nonoperating
3.48 g
RMS
, 5 Hz to 500 Hz, 10 minutes each axis
Shock, Amplifiers
50 g, 11 ms duration, half-sine pulses
Electro-Magnetic Compliance
Meets FCC Part 15, Subpart B, Class A
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......