TCPA300 and TCPA400 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 15
Table 5- 10: Maximum current ratings for TCPA300-compatible probes
Probe model
Range
Maximum
current rating,
Amps
Current loop
used for test
1
Current source
output, Amps
TCP312
10 A/V
±
30
5 T
±
6
TCP305
10 A/V
±
50
5 T
±
10
TCP303
50 A/V
±
212
2
50 T
±
4.24
1
Refer to page 5- 4 for instructions on making the current loops.
2
The TCP303 is a 150 A RMS probe with a peak current of 212 A. Apply current for
only a short amount of time.
5.
Slowly increase the output of the current source until the OVERLOAD LED
on the amplifier is on.
CAUTION.
Do not leave the current turned up longer than necessary -- damage to
the equipment or personal injury may result.
6.
Decrease the output of the current source to zero.
7.
Record the results (pass/fail) on the test record.
8.
Repeat steps 3 through 7 for the opposite polarity.
This completes the performance verification for the amplifiers.
If the TCPA300 or TCPA400 amplifier fails any of the verification tests, refer to
the
Adjustments
section for servicing information.
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......