Disassembly
7- 10
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
Replacing the Power Supply
To replace the power supply, refer to Figure 7--7 and do the following:
1.
Disconnect the AC and DC power connectors from the power supply.
2.
Remove the four screws from the power supply.
3.
Remove the power supply from the chassis.
4.
Do in reverse steps 3 through 1 to install a new power supply.
AC connector
DC connector
Screws (4)
Torque screws to 8 in-lbs
when reassembling
Figure 7- 7: Replacing the power supply