TCP305 and TCP312 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 27
Make or verify the equipment settings in Table 5--15.
Table 5- 15: Equipment settings for bandwidth
Oscilloscope
Vertical input impedance
50
Ω
Time base
200 ns/division
Record length
500
Coupling
DC
Offset
0 V (mid-scale)
Trigger type
Edge
Trigger mode
Auto
Trigger position
50%
Acquisition mode
Average
Number of waveforms to average
8
Measurement type
Peak-to-Peak
Leveled sine wave generator
Frequency
3 MHz
Amplitude
3 V
p-p
TCPA300
Coupling
DC
1.
Do not clamp the current probe around any conductor, but make sure the
jaws are locked shut.
2.
Press the amplifier
PROBE DEGAUSS AUTOBALANCE
button. Wait for
the degauss/autobalance routine to complete before proceeding. The routine
is complete when the indicator light turns green.
3.
Clamp the current probe around the HF current loop.
4.
Enable the output of the leveled sinewave generator.
5.
Verify the output level is what is listed for the probe in Table 5--16.
6.
Using the peak-peak measurement capability of the oscilloscope, measure
the peak-peak reading and record it in Table 5--16 as
M
1
.
Equipment Settings
Procedure
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......